TY - JOUR
T1 - Percutaneous femoral implantation of an arterial port catheter for intraarterial chemotherapy
T2 - Feasibility and predictive factors of long-term functionality
AU - Deschamps, Frederic
AU - Rao, Pramod
AU - Teriitehau, Christophe
AU - Hakime, Antoine
AU - Malka, David
AU - Boige, Valerie
AU - Ducreux, Michel
AU - Elias, Dominique
AU - Goere, Diane
AU - De Baere, Thierry
PY - 2010/11/1
Y1 - 2010/11/1
N2 - Purpose To evaluate the feasibility, functionality, and dysfunctions of an arterial port catheter implanted via the femoral artery. Materials and Methods From November 2001 to May 2008, 93 consecutive patients (mean age 57 years old) with unresectable hepatic colorectal metastases were referred for intraarterial chemotherapy. The arterial port catheters were percutaneously implanted via the femoral artery. The catheter tips were placed as "free-floating" in the common hepatic artery (technique 1), "fixed" in the gastroduodenal artery (technique 2), or inserted in a segmental hepatic artery (technique 3). Embolization of the right gastric artery was always attempted. Results The technical success rate of the femoral approach was 94% (n = 88 of 93). Intraarterial chemotherapy (average 7.3 courses) was administered to 84 patients. Migration and occlusion of the catheters occurred in 12% (n = 10 of 84) and 11% (n = 9 of 84) of patients, and extrahepatic perfusion occurred in 30% (n = 25 of 84) of patients. Catheter migration occurred significantly more frequently with technique 1 (50%; n = 3 of 6) than with technique 2 (11%; n = 7 of 64; P = .03) or technique 3 (0%; n = 0 of 14; P = .02). Occurrence of gastroduodenal ulcerations was significantly lower (P = .01) when embolization of the right gastric artery was performed (8%; n = 4 of 48) than when it was not (28%; n = 11 of 36). The success rate of embolization of the right gastric artery significantly improved (P = .006) from the first half of patients treated to the second half, resulting in a significant (P = .02) decrease in the occurrence of ulcerations from 28% (n = 12 of 42) in the first half of patients treated to 7% (n = 3 of 42) in the second half. Conclusions Percutaneous femoral placement of an arterial port catheter is highly feasible. Right gastric artery embolization and use of techniques 2 and 3 are good predictive factors for long-term functionality.
AB - Purpose To evaluate the feasibility, functionality, and dysfunctions of an arterial port catheter implanted via the femoral artery. Materials and Methods From November 2001 to May 2008, 93 consecutive patients (mean age 57 years old) with unresectable hepatic colorectal metastases were referred for intraarterial chemotherapy. The arterial port catheters were percutaneously implanted via the femoral artery. The catheter tips were placed as "free-floating" in the common hepatic artery (technique 1), "fixed" in the gastroduodenal artery (technique 2), or inserted in a segmental hepatic artery (technique 3). Embolization of the right gastric artery was always attempted. Results The technical success rate of the femoral approach was 94% (n = 88 of 93). Intraarterial chemotherapy (average 7.3 courses) was administered to 84 patients. Migration and occlusion of the catheters occurred in 12% (n = 10 of 84) and 11% (n = 9 of 84) of patients, and extrahepatic perfusion occurred in 30% (n = 25 of 84) of patients. Catheter migration occurred significantly more frequently with technique 1 (50%; n = 3 of 6) than with technique 2 (11%; n = 7 of 64; P = .03) or technique 3 (0%; n = 0 of 14; P = .02). Occurrence of gastroduodenal ulcerations was significantly lower (P = .01) when embolization of the right gastric artery was performed (8%; n = 4 of 48) than when it was not (28%; n = 11 of 36). The success rate of embolization of the right gastric artery significantly improved (P = .006) from the first half of patients treated to the second half, resulting in a significant (P = .02) decrease in the occurrence of ulcerations from 28% (n = 12 of 42) in the first half of patients treated to 7% (n = 3 of 42) in the second half. Conclusions Percutaneous femoral placement of an arterial port catheter is highly feasible. Right gastric artery embolization and use of techniques 2 and 3 are good predictive factors for long-term functionality.
UR - http://www.scopus.com/inward/record.url?scp=77958520374&partnerID=8YFLogxK
U2 - 10.1016/j.jvir.2010.08.003
DO - 10.1016/j.jvir.2010.08.003
M3 - Article
C2 - 21029950
AN - SCOPUS:77958520374
SN - 1051-0443
VL - 21
SP - 1681
EP - 1688
JO - Journal of Vascular and Interventional Radiology
JF - Journal of Vascular and Interventional Radiology
IS - 11
ER -