TY - JOUR
T1 - Radiofrequency liver ablation
T2 - Experimental comparative study of water-cooled versus expandable systems
AU - De Baere, T.
AU - Denys, A.
AU - Wood, B. J.
AU - Lassau, N.
AU - Kardache, M.
AU - Vilgrain, V.
AU - Menu, Y.
AU - Roche, A.
PY - 2001/1/1
Y1 - 2001/1/1
N2 - OBJECTIVE. We evaluate the uniformity and reproducibility of thermal lesion ablation and quantify the volume of tissue destruction and hemorrhage induced with two different commercially available radiofrequency ablation devices. MATERIALS AND METHODS. A four-array anchor expandable needle electrode and a triple-cluster cooled-tip needle electrode were used to induce lesions in three explanted calf livers and in vivo in eight swine livers. The sizes of the radiofrequency-induced lesions were macroscopically evaluated by measuring two perpendicular dimensions immediately after the experiment. Bleeding was evaluated by weighing gauze swabs used to dry the hemorrhage caused by electrode insertions. RESULTS. In explanted liver, the mean diameter of the radiofrequency-induced lesion was 5.3 ± 0.7 cm for the cooled-tip needle and 3.7 ± 0.4 cm for the expandable needle (p = 0.042), which correspond to approximate volumes of 65.35 ± 26.22 cm3 and 26.67 ± 9.59 cm3, respectively (p < 0.002). In vivo, the mean diameter was 3.7 ± 0.4 cm for the cooled-tip needle and 3 ± 0.4 cm for the expandable needle (p < 0.0001), which correspond to approximate volumes of 24.18±7.56 cm3 and 11.16 ± 3.65 cm3, respectively (p < 0.0001). Blood loss attained a median value of 3.5 g for the cooled-tip needle and 2.6 g for the expandable needle; this difference was not statistically significant (p = 0.06). CONCLUSION. The cooled-tip needle induced significantly larger lesions than the expandable needle, but the lesions produced by the expandable needle are more reproducible, uniform, and spheric. The larger size of the lesions produced by the cooled-tip needle may be attributed to the higher maximum power used by the generator and the higher energy deposition, which is due to the cooling of the needle electrode.
AB - OBJECTIVE. We evaluate the uniformity and reproducibility of thermal lesion ablation and quantify the volume of tissue destruction and hemorrhage induced with two different commercially available radiofrequency ablation devices. MATERIALS AND METHODS. A four-array anchor expandable needle electrode and a triple-cluster cooled-tip needle electrode were used to induce lesions in three explanted calf livers and in vivo in eight swine livers. The sizes of the radiofrequency-induced lesions were macroscopically evaluated by measuring two perpendicular dimensions immediately after the experiment. Bleeding was evaluated by weighing gauze swabs used to dry the hemorrhage caused by electrode insertions. RESULTS. In explanted liver, the mean diameter of the radiofrequency-induced lesion was 5.3 ± 0.7 cm for the cooled-tip needle and 3.7 ± 0.4 cm for the expandable needle (p = 0.042), which correspond to approximate volumes of 65.35 ± 26.22 cm3 and 26.67 ± 9.59 cm3, respectively (p < 0.002). In vivo, the mean diameter was 3.7 ± 0.4 cm for the cooled-tip needle and 3 ± 0.4 cm for the expandable needle (p < 0.0001), which correspond to approximate volumes of 24.18±7.56 cm3 and 11.16 ± 3.65 cm3, respectively (p < 0.0001). Blood loss attained a median value of 3.5 g for the cooled-tip needle and 2.6 g for the expandable needle; this difference was not statistically significant (p = 0.06). CONCLUSION. The cooled-tip needle induced significantly larger lesions than the expandable needle, but the lesions produced by the expandable needle are more reproducible, uniform, and spheric. The larger size of the lesions produced by the cooled-tip needle may be attributed to the higher maximum power used by the generator and the higher energy deposition, which is due to the cooling of the needle electrode.
UR - http://www.scopus.com/inward/record.url?scp=0035183913&partnerID=8YFLogxK
U2 - 10.2214/ajr.176.1.1760187
DO - 10.2214/ajr.176.1.1760187
M3 - Article
C2 - 11133564
AN - SCOPUS:0035183913
SN - 0361-803X
VL - 176
SP - 187
EP - 192
JO - American Journal of Roentgenology
JF - American Journal of Roentgenology
IS - 1
ER -