TY - JOUR
T1 - Safety and Efficacy of Percutaneous Liver Microwave Ablation Using a Fully Water-Cooled Choke Ring Antenna
T2 - First Multicenter Clinical Report
AU - Blain, Maxime
AU - Narayanan, Govindarajan
AU - Ricoeur, Alexis
AU - Kobe, Adrian
AU - Mahendra, Ashwin M.
AU - Jacks, Blake
AU - Letty, Quentin
AU - Bonnet, Baptiste
AU - Tselikas, Lambros
AU - Deschamps, Frederic
AU - de Baère, Thierry
N1 - Publisher Copyright:
© 2023, The Author(s).
PY - 2023/8/1
Y1 - 2023/8/1
N2 - Introduction: The safety and efficacy of a microwave ablation (MWA) system for the liver with novel technologies in field control, antenna cooling through the inner part of the choke ring, and dual temperature monitoring were evaluated in this multicenter retrospective study. Material and Methods: Ablation characteristics and efficacy were assessed on follow-up imaging (computed tomography or magnetic resonance imaging). Safety was evaluated according to CTCAE classification. Results: Eighty-seven liver tumors (65 metastases and 22 hepatocellular carcinomas) measuring 17.8 ± 7.9 mm were treated in 68 patients. Ablation zones measured 35.6 ± 11 mm in longest diameter. The coefficients of variation of the longest and shortest ablation diameters were 30.1% and 26.4%, respectively. The mean sphericity index of the ablation zone was 0.78 ± 0.14. Seventy-one ablations (82%) had a sphericity index above 0.66. At 1 month, all tumors demonstrated complete ablation with margins of 0–5 mm, 5–10 mm, and greater than 10 mm achieved in 22%, 46%, and 31% of tumors, respectively. After a median follow-up of 10 months, local tumor control was achieved in 84.7% of treated tumors after a single ablation and in 86% after one patient received a second ablation. One grade 3 complication (stress ulcer) occurred, but was unrelated to the procedure. Ablation zone size and geometry in this clinical study were in accordance with previously reported in vivo preclinical findings. Conclusion: Promising results were reported for this MWA device. The high spherical index, reproducibility, and predictability of the resulting treatment zones translated to a high percentage of adequate safety margins, providing good local control rate.
AB - Introduction: The safety and efficacy of a microwave ablation (MWA) system for the liver with novel technologies in field control, antenna cooling through the inner part of the choke ring, and dual temperature monitoring were evaluated in this multicenter retrospective study. Material and Methods: Ablation characteristics and efficacy were assessed on follow-up imaging (computed tomography or magnetic resonance imaging). Safety was evaluated according to CTCAE classification. Results: Eighty-seven liver tumors (65 metastases and 22 hepatocellular carcinomas) measuring 17.8 ± 7.9 mm were treated in 68 patients. Ablation zones measured 35.6 ± 11 mm in longest diameter. The coefficients of variation of the longest and shortest ablation diameters were 30.1% and 26.4%, respectively. The mean sphericity index of the ablation zone was 0.78 ± 0.14. Seventy-one ablations (82%) had a sphericity index above 0.66. At 1 month, all tumors demonstrated complete ablation with margins of 0–5 mm, 5–10 mm, and greater than 10 mm achieved in 22%, 46%, and 31% of tumors, respectively. After a median follow-up of 10 months, local tumor control was achieved in 84.7% of treated tumors after a single ablation and in 86% after one patient received a second ablation. One grade 3 complication (stress ulcer) occurred, but was unrelated to the procedure. Ablation zone size and geometry in this clinical study were in accordance with previously reported in vivo preclinical findings. Conclusion: Promising results were reported for this MWA device. The high spherical index, reproducibility, and predictability of the resulting treatment zones translated to a high percentage of adequate safety margins, providing good local control rate.
KW - Interventional oncology
KW - Liver neoplasms
KW - Microwave ablation
UR - http://www.scopus.com/inward/record.url?scp=85164348675&partnerID=8YFLogxK
U2 - 10.1007/s00270-023-03481-3
DO - 10.1007/s00270-023-03481-3
M3 - Article
C2 - 37430015
AN - SCOPUS:85164348675
SN - 0174-1551
VL - 46
SP - 1086
EP - 1091
JO - CardioVascular and Interventional Radiology
JF - CardioVascular and Interventional Radiology
IS - 8
ER -