Safety and Efficacy of Percutaneous Liver Microwave Ablation Using a Fully Water-Cooled Choke Ring Antenna: First Multicenter Clinical Report

Maxime Blain, Govindarajan Narayanan, Alexis Ricoeur, Adrian Kobe, Ashwin M. Mahendra, Blake Jacks, Quentin Letty, Baptiste Bonnet, Lambros Tselikas, Frederic Deschamps, Thierry de Baère

    Résultats de recherche: Contribution à un journalArticleRevue par des pairs

    2 Citations (Scopus)

    Résumé

    Introduction: The safety and efficacy of a microwave ablation (MWA) system for the liver with novel technologies in field control, antenna cooling through the inner part of the choke ring, and dual temperature monitoring were evaluated in this multicenter retrospective study. Material and Methods: Ablation characteristics and efficacy were assessed on follow-up imaging (computed tomography or magnetic resonance imaging). Safety was evaluated according to CTCAE classification. Results: Eighty-seven liver tumors (65 metastases and 22 hepatocellular carcinomas) measuring 17.8 ± 7.9 mm were treated in 68 patients. Ablation zones measured 35.6 ± 11 mm in longest diameter. The coefficients of variation of the longest and shortest ablation diameters were 30.1% and 26.4%, respectively. The mean sphericity index of the ablation zone was 0.78 ± 0.14. Seventy-one ablations (82%) had a sphericity index above 0.66. At 1 month, all tumors demonstrated complete ablation with margins of 0–5 mm, 5–10 mm, and greater than 10 mm achieved in 22%, 46%, and 31% of tumors, respectively. After a median follow-up of 10 months, local tumor control was achieved in 84.7% of treated tumors after a single ablation and in 86% after one patient received a second ablation. One grade 3 complication (stress ulcer) occurred, but was unrelated to the procedure. Ablation zone size and geometry in this clinical study were in accordance with previously reported in vivo preclinical findings. Conclusion: Promising results were reported for this MWA device. The high spherical index, reproducibility, and predictability of the resulting treatment zones translated to a high percentage of adequate safety margins, providing good local control rate.

    langue originaleAnglais
    Pages (de - à)1086-1091
    Nombre de pages6
    journalCardioVascular and Interventional Radiology
    Volume46
    Numéro de publication8
    Les DOIs
    étatPublié - 1 août 2023

    Contient cette citation