T1 - Toward a Standardization of Ultrasound Scanners for Dynamic Contrast-Enhanced Ultrasonography
T2 - Methodology and Phantoms
AU - Pitre-Champagnat, Stéphanie
AU - Coiffier, Bénédicte
AU - Jourdain, Laurène
AU - Benatsou, Baya
AU - Leguerney, Ingrid
AU - Lassau, Nathalie
N1 - Publisher Copyright:
© 2017 World Federation for Ultrasound in Medicine & Biology
PY - 2017/11/1
Y1 - 2017/11/1
N2 - The standardization of ultrasound scanners for dynamic contrast-enhanced ultrasonography (DCE-US) is mandatory for evaluation of clinical multicenter studies. We propose a robust method using a phantom for measuring the variation of the harmonic signal intensity obtained from the area under the time-intensity curve versus various contrast-agent concentrations. The slope of this measured curve is the calibration parameter. We tested our method on two devices from the same manufacturer (AplioXV and Aplio500, Toshiba, Tokyo, Japan) using the same settings as defined for a French multicenter study. The Aplio500's settings were adjusted to match the slopes of the AplioXV, resulting in the following settings on the Aplio500: at 3.5 MHz: MI = 0.15; CG = 35 dB and at 8 MHz: MI = 0.10; CG = 32 dB. This calibration method is very important for future DCE-US multicenter studies.
AB - The standardization of ultrasound scanners for dynamic contrast-enhanced ultrasonography (DCE-US) is mandatory for evaluation of clinical multicenter studies. We propose a robust method using a phantom for measuring the variation of the harmonic signal intensity obtained from the area under the time-intensity curve versus various contrast-agent concentrations. The slope of this measured curve is the calibration parameter. We tested our method on two devices from the same manufacturer (AplioXV and Aplio500, Toshiba, Tokyo, Japan) using the same settings as defined for a French multicenter study. The Aplio500's settings were adjusted to match the slopes of the AplioXV, resulting in the following settings on the Aplio500: at 3.5 MHz: MI = 0.15; CG = 35 dB and at 8 MHz: MI = 0.10; CG = 32 dB. This calibration method is very important for future DCE-US multicenter studies.
KW - Dynamic contrast-enhanced ultrasound (DCE-US)
KW - Flow phantom
KW - Methodologies of ultrasound calibration
UR - http://www.scopus.com/inward/record.url?scp=85026651759&partnerID=8YFLogxK
U2 - 10.1016/j.ultrasmedbio.2017.06.032
DO - 10.1016/j.ultrasmedbio.2017.06.032
M3 - Article
C2 - 28779957
AN - SCOPUS:85026651759
SN - 0301-5629
VL - 43
SP - 2670
EP - 2677
JO - Ultrasound in Medicine and Biology
JF - Ultrasound in Medicine and Biology
IS - 11
ER -